Tipping

Tipping

The tip is the first thing that touches tissue. Tipping reshapes the distal end into an atraumatic profile and, where required, closes the transition between an inner liner, a braid or coil layer and the outer jacket so there is no step for tissue or a guidewire to catch on.

Tip geometries

  • Tapered and radiused atraumatic tips
  • Guidewire-compatible tapers matched to a target wire diameter
  • Soft distal sections formed from a lower-durometer outer layer

Process notes

  • Tip length, taper angle and wall transition set from the drawing
  • Inner lumen supported during forming so the tip does not neck down
  • Reinforcement layers terminated and encapsulated so no braid ends are exposed

Typically combined with

Ultra-soft TPU or Pebax outer layers on the distal section — the same approach used on neurovascular catheters, where a softer tip reduces trauma to the vessel wall.

Contact Us

Comprehensive medical device manufacturing solutions